874-VQL: Difference between revisions

From GRWiki
Jump to navigation Jump to search
mNo edit summary
No edit summary
 
Line 14: Line 14:
{{Catalog History}}
{{Catalog History}}
}}
}}
The {{Title|General Radio 874-VQL}} is a "voltmeter detector" with two [[GR-874 connector]]s for the signal, and another GR-874 for the rectified output.  It was introduced in {{Catalog R}} and remained available through {{Catalog 1973}}, after which only the 874-VQL version with locking connectors remained in the catalog.  
The {{Title|General Radio 874-VQL}} is a "voltmeter detector" with two locking [[GR-874 connector]]s for the signal, and a non-locking GR-874 for the rectified output.  It was introduced in {{Catalog R}} and remained available through {{Catalog 1978}}.


The 874-VQL is used in frequency-response measurements with a [[1750-A|1750-A Sweep Drive]] and [[1263-A|1263-A Amplitude Regulating Power Supply]].
The 874-VQL is used in frequency-response measurements with a [[1750-A|1750-A Sweep Drive]] and [[1263-A|1263-A Amplitude Regulating Power Supply]].
Line 29: Line 29:
==Links==
==Links==


* [[874-VQ]] − Version with non-locking connectors
* [[Media:GR Exp 874 Locking 10_1961.pdf|Experimenter October 1961 describing 874 Locking Connectors]]
* [[Media:GR Exp 874 Locking 10_1961.pdf|Experimenter October 1961 describing 874 Locking Connectors]]
* [[Media:GR Exp 874 Locking 07_1962.pdf|Experimenter July 1962 describing 874 Locking Connectors]]
* [[Media:GR Exp 874 Locking 07_1962.pdf|Experimenter July 1962 describing 874 Locking Connectors]]

Latest revision as of 15:05, 20 June 2024

General Radio 874-VQL
Voltmeter detector
General Radio 874-VQL

Available from 1963 to 1973

Manuals
Catalog History
 Document Year Page
Catalog R 1963 67
Catalog S 1965 86
Catalog T 1968 143
Catalog73 1973 249
Catalog78 1978 101
(All manuals in PDF format unless noted otherwise)

The General Radio 874-VQL is a "voltmeter detector" with two locking GR-874 connectors for the signal, and a non-locking GR-874 for the rectified output. It was introduced in Catalog R (1963) and remained available through Catalog 1978.

The 874-VQL is used in frequency-response measurements with a 1750-A Sweep Drive and 1263-A Amplitude Regulating Power Supply. An 874-VI Voltmeter Indicator can be used for calibrated measurements from 0.1 to 2 V.

The 874-VQL contains a 1N23B diode and a mica capacitor. It is similar to the 874-VR Voltmeter Rectifier, but omits the latter's 50 Ω series resistor, and includes matching elements to compensate for the discontinuity caused by the diode capacitance.

Key Specifications

  • Frequency range: 500 kHz to 2 GHz ("in limited application" up to 7 GHz)
  • Maximum input: 2 V
  • VSWR: ≤1.1 @ 1 GHz, ≤1.2 @ 2 GHz; rising rapidly above 2.5 GHz
  • Resonant frequency: 3.6 GHz

Links

Photos