821-A: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
Line 16: Line 16:
The {{Title|General Radio 821-A Twin-T Impedance Measuring Circuit}} were introduced in {{Catalog K2}} and remained available through {{Catalog O}}.  
The {{Title|General Radio 821-A Twin-T Impedance Measuring Circuit}} were introduced in {{Catalog K2}} and remained available through {{Catalog O}}.  


The Type 821-A is an impedance measuring device operating between 460 kHz and 40 Mhz. Measurements are made using the Parallel-T null circuit and the parallel substitution method. Its calibrated in capacitance and conductance. Just some of the applications are the measurement of capacitance and dissipation factor of capacitors, and the inductance and Q of coils. The 821-A requires a generator and detector to accomplish a null. An oscillator of sufficient output at the frequency of interest, the Type [[1330-A]] Bridge Oscillator is the recommended generator. A radio receiver equipped with a BFO and S-Meter makes a suitable detector. A well shielded cable with a coaxial connection to the receiver is highly recommended.
The Type 821-A is an impedance measuring device operating between 460 kHz and 40 Mhz. Measurements are made using the Parallel-T null circuit and the parallel substitution method. Its calibrated in capacitance and conductance. Just some of the applications are the measurement of capacitance and dissipation factor of capacitors, and the inductance and Q of coils. The 821-A requires a generator and detector to accomplish a null. An oscillator of sufficient output at the frequency of interest, the Type [[1330-A]] Bridge Oscillator is the recommended generator. A communications type radio receiver equipped with a BFO and RF gain control makes a suitable detector. A well shielded cable with a coaxial connection to the receiver is highly recommended.


The Type 821-A is direct reading instrument for capacitance and conductance, some math is required to obtain other impedances.
The Type 821-A is direct reading instrument for capacitance and conductance, some math is required to obtain other impedances.

Revision as of 20:46, 24 May 2024

General Radio 821-A
Twin-T Impedance Measuring Circuit
General Radio 821-A Twin-T Impedance Measuring Circuit

Available from 1942 to 1956

Manuals
Catalog History
 Document Year Page
Catalog K2 1942 70
Catalog K3 1944 68
Catalog K4 1945 68
Catalog L 1948 72
Catalog M 1951 82
Catalog N 1954 98
Catalog O 1956 42
(All manuals in PDF format unless noted otherwise)

The General Radio 821-A Twin-T Impedance Measuring Circuit were introduced in Catalog K2 (1942) and remained available through Catalog O (1956).

The Type 821-A is an impedance measuring device operating between 460 kHz and 40 Mhz. Measurements are made using the Parallel-T null circuit and the parallel substitution method. Its calibrated in capacitance and conductance. Just some of the applications are the measurement of capacitance and dissipation factor of capacitors, and the inductance and Q of coils. The 821-A requires a generator and detector to accomplish a null. An oscillator of sufficient output at the frequency of interest, the Type 1330-A Bridge Oscillator is the recommended generator. A communications type radio receiver equipped with a BFO and RF gain control makes a suitable detector. A well shielded cable with a coaxial connection to the receiver is highly recommended.

The Type 821-A is direct reading instrument for capacitance and conductance, some math is required to obtain other impedances.

Specifications

  • Frequency Range: 460 kHz to 40 MHz
  • Conductance Range:
  • 0-100 μS at 1 MHz
  • 0-300 μS at 3 MHz
  • 0-1000 μS at 10 MHz
  • 0-3000 μS at 30 MHz
  • Accuracy Capacitance: ±1 pF ±0.1%
  • Accuracy Conductance: ±0.1% of full scale ±2% of actual dial reading

Links

Photos