1604-B: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(12 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{GR Product
{{GR Product
|model=1604-B
|model=1604-B
|codes=FATTY
|class=impedance measurement
|class=impedance measurement
|summary=comparison bridge
|summary=Comparison bridge
|image=GR 1604-B Front.jpg
|image=GR 1604-B Front.jpg
|caption=General Radio 1604-B Comparison Bridge
|caption=General Radio 1604-B Comparison Bridge
Line 11: Line 12:
|manuals=
|manuals=
* [[Media:GR 1604-B Impedance Comparison Bridge 790-B 1956.pdf|General Radio 1604-B Impedance Comparison Bridge 790-B 1956]]
* [[Media:GR 1604-B Impedance Comparison Bridge 790-B 1956.pdf|General Radio 1604-B Impedance Comparison Bridge 790-B 1956]]
{{Catalog History}}
}}
}}
The '''General Radio 1604-B Comparison Bridge''' was introduced in 1954 Catalog N and remained available through Catalog O 1956. The '''General Radio 1604-B Comparison Bridge''' was introduced in 1954 Catalog N and remained available through Catalog O 1956.
The {{Title|General Radio 1604-B Comparison Bridge}} was introduced in {{Catalog N}} and remained available through {{Catalog O}}.
The Type 1604-B was designed for sorting, adjusting and comparing two devices at three different frequencies.  Impedance difference is measured in two ranges 5% and 20% using front panel switch, using a built generator and detector equiped with a 2AP1 CRT.  A balance is achieved when the vertical deflection is at a minimum. Disipation factor difference is read directly from the dial, one range.  The main change from the [[1604-A]] is the addition of the 400 Hz test frequency.


It was designed for sorting, adjusting and comparing two devices at three different frequencies. 
Impedance difference can be measured in two ranges, 5% or 20%, selected by a front panel switch, using a built-in generator and detector equipped with a 2AP1 CRT. 
Balance is achieved when the vertical deflection is at a minimum.
Dissipation factor difference is read directly from the dial in one range.
The main change from the [[1604-A]] is the addition of a 400 Hz test frequency.


==Specifications==
==Specifications==
* '''Deviation Range:''' Impedance difference, ±5% and ±20%, switched. Disipation factor difference, ±.006 at 400 Hz. ±.015 at 1 kHz. ±.075 at 5 kHz.
* '''Deviation Range:''' Impedance difference, ±5% and ±20%, switched. Dissipation factor difference, ±0.006 at 400 Hz, ±0.015 at 1 kHz, ±0.075 at 5 kHz
* '''Test Frequency:''' 400 Hz, 1 kHz, and 5 kHz; The frequency is within ±3% of the nominal value.
* '''Test Frequency:''' 400 Hz, 1 kHz, or 5 kHz within ±3%
* '''Voltage Applied to Unknown:''' Approximately one volt, for impedances above 500 Ω. For lower values of impedance the voltage is decreased.
* '''Voltage Applied to Unknown:''' Approximately 1 V for impedances above 500 Ω, decreasing for lower values of impedance


==Links==
==Links==
* [[Media:GRwiki Exp 1604-A 12_1952.pdf|Experimenter describing Type 1604-A December 1952]]
* [[Media:GRwiki Exp 1604-A 12_1952.pdf|Experimenter December 1952 describing Type 1604-A]]
* [[Media:GRwiki Exp 1604-B 02_1954.pdf|Experimenter describing Type 1604-B February 1954]]
* [[Media:GRwiki Exp 1604-B 02_1954.pdf|Experimenter February 1954 describing Type 1604-B]]
* Predecessor: [[1604-A]] (1952)
* Successor: [[1605-A|General Radio 1605-A Impedance Comparator]] (1956)


==Photos==
==Photos==

Latest revision as of 15:11, 25 April 2024

General Radio 1604-B
Comparison bridge
General Radio 1604-B Comparison Bridge

Available from 1954 to 1956

Manuals
Catalog History
 Document Year Page
Catalog N 1954 93
Catalog O 1956 31
(All manuals in PDF format unless noted otherwise)

The General Radio 1604-B Comparison Bridge was introduced in Catalog N (1954) and remained available through Catalog O (1956).

It was designed for sorting, adjusting and comparing two devices at three different frequencies.

Impedance difference can be measured in two ranges, 5% or 20%, selected by a front panel switch, using a built-in generator and detector equipped with a 2AP1 CRT. Balance is achieved when the vertical deflection is at a minimum.

Dissipation factor difference is read directly from the dial in one range. The main change from the 1604-A is the addition of a 400 Hz test frequency.

Specifications

  • Deviation Range: Impedance difference, ±5% and ±20%, switched. Dissipation factor difference, ±0.006 at 400 Hz, ±0.015 at 1 kHz, ±0.075 at 5 kHz
  • Test Frequency: 400 Hz, 1 kHz, or 5 kHz within ±3%
  • Voltage Applied to Unknown: Approximately 1 V for impedances above 500 Ω, decreasing for lower values of impedance

Links

Photos