1230-A: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
 
(7 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{GR Product
{{GR Product
|model=1230-A
|model=1230-A
|codes=MASON,1230-9701
|class=meters
|class=meters
|summary=electrometer and dc amplifier
|summary=Electrometer and DC amplifier
|image=GR 1230-A Front Close-up.jpg
|image=GR 1230-A Front Close-up.jpg
|caption=General Radio 1230-A Electrometer and DC Amplifier
|caption=General Radio 1230-A Electrometer and DC Amplifier
Line 11: Line 12:
|manuals=
|manuals=
* [[Media:GRwiki 1230A Electometer 1230-0100-F 02_1962.pdf|General Radio 1230-A Electrometer and DC Amplifier Manual 1230-0100F 1962]]
* [[Media:GRwiki 1230A Electometer 1230-0100-F 02_1962.pdf|General Radio 1230-A Electrometer and DC Amplifier Manual 1230-0100F 1962]]
{{Catalog History}}
}}
}}
The '''General Radio 1230-A Electrometer and DC Amplifier''' was introduced in 1956 Catalog O and remained available through Catalog T 1968.  
The {{Title|General Radio 1230-A Electrometer and DC Amplifier}} was introduced in {{Catalog O}} and remained available through {{Catalog T}}.  
 


The Type 1230-A is laboratory instrument designed for the measurement of extremely small voltages or currents and conversely extremely large resistances. It achieves this with the [[tube:5886|5886 electrometer vacuum tube]] used in the first stage of amplification.  The combined gain of the three stages result in a transconductance in the millions of μS. The output of the amplifier is read from the panel meter, or an optional recorder connection at the rear panel. The recorder terminals are in series with the meter, so a resistor is required without a recorder connected.


==Specifications==
==Specifications==


* '''RANGES OF MEASUREMENT'''
* '''RANGES OF MEASUREMENT'''
* '''Voltage:''' ±30, 100, and 300 mV; ±1, 3, and 10 V; dc full scale.
* '''Voltage:''' ±30, 100, and 300 mV, ±1, 3, and 10 V DC full scale
* '''Current:''' ±1 mA dc, full scale, to ±0.3 pA full scale; Measured in terms of voltage.
* '''Current:''' ±1 mA DC full scale, to ±0.3 pA full scale; Measured in terms of voltage
* '''Resistance:''' direct reading from 300  kΩ to 10 TΩ full scale; There are 16 ranges, two per decade. Voltage across the unknown resistance is 9.1 V.
* '''Resistance:''' direct reading from 300  kΩ to 10 TΩ full scale in 16 ranges, two per decade. Voltage across the unknown resistance is 9.1 V
* '''ACCURACY'''
* '''ACCURACY'''
* '''Voltage:''' ±2% of full scale on the five highest ranges, ±4% of full scale on the 30 mV range.
* '''Voltage:''' ±2% of full scale on the five highest ranges, ±4% of full scale on the 30 mV range
* '''Current:''' ±3% of full scale from 1 mA to 1 nA, ±10% of full scale from 300 pA to 0.3 pA.
* '''Current:''' ±3% of full scale from 1 mA to 1 nA, ±10% of full scale from 300 pA to 0.3 pA
* '''Resistance:''' ±3% from 300 kΩ to 10 MΩ at full scale (low-resistance end), ±8% from 3 MΩ to 10 TΩ.
* '''Resistance:''' ±3% from 300 kΩ to 10 MΩ at full scale (low-resistance end), ±8% from 3 MΩ to 10 TΩ


==Links==
==Links==


* [[Media:GR Exp 1230-A 03_1956.pdf|Experimenter describing Type 1230-A March 1956]]
* [[Media:GR Exp 1230-A 03_1956.pdf|Experimenter March 1956 describing Type 1230-A]]
* [[Media:Tung-Sol 6418 Tubedata 1963.pdf|Tung-Sol 6418 Datasheet]]
* [[Media:Tung-Sol 6418 Tubedata 1963.pdf|Tung-Sol 6418 Datasheet]]
* [[Media:Tung-Sol 5886 Tubedata 1963.pdf|Tung-Sol 5886 Datasheet]]
* [[Media:Tung-Sol 5886 Tubedata 1963.pdf|Tung-Sol 5886 Datasheet]]
Line 43: Line 44:
GR 1230-A Bottom View.jpg
GR 1230-A Bottom View.jpg
GR 1230-A V1 V2 Close-up.jpg
GR 1230-A V1 V2 Close-up.jpg
</gallery>
</gallery>


[[Category:Meters]]
[[Category:Meters]]
[[Category:Detectors Amplifiers]]
[[Category:Detectors Amplifiers]]

Latest revision as of 11:03, 30 April 2024

General Radio 1230-A
Electrometer and DC amplifier
General Radio 1230-A Electrometer and DC Amplifier

Available from 1956 to 1968

Manuals
Catalog History
 Document Year Page
Catalog O 1956 2
Catalog P 1959 130
Catalog Q 1961 128
Catalog R 1963 140
Catalog S 1965 174
Catalog T 1968 239
(All manuals in PDF format unless noted otherwise)

The General Radio 1230-A Electrometer and DC Amplifier was introduced in Catalog O (1956) and remained available through Catalog T (1968).

The Type 1230-A is laboratory instrument designed for the measurement of extremely small voltages or currents and conversely extremely large resistances. It achieves this with the 5886 electrometer vacuum tube used in the first stage of amplification. The combined gain of the three stages result in a transconductance in the millions of μS. The output of the amplifier is read from the panel meter, or an optional recorder connection at the rear panel. The recorder terminals are in series with the meter, so a resistor is required without a recorder connected.

Specifications

  • RANGES OF MEASUREMENT
  • Voltage: ±30, 100, and 300 mV, ±1, 3, and 10 V DC full scale
  • Current: ±1 mA DC full scale, to ±0.3 pA full scale; Measured in terms of voltage
  • Resistance: direct reading from 300  kΩ to 10 TΩ full scale in 16 ranges, two per decade. Voltage across the unknown resistance is 9.1 V
  • ACCURACY
  • Voltage: ±2% of full scale on the five highest ranges, ±4% of full scale on the 30 mV range
  • Current: ±3% of full scale from 1 mA to 1 nA, ±10% of full scale from 300 pA to 0.3 pA
  • Resistance: ±3% from 300 kΩ to 10 MΩ at full scale (low-resistance end), ±8% from 3 MΩ to 10 TΩ

Links

Photos