Revision as of 02:18, 4 April 2024

General Radio 1657
digital RLC meter
Genrad 1657 Digibridge

Available from 1976 to (?)

Manuals
Catalog History
 Document Year Page
Catalog78 1978 22
(All manuals in PDF format unless noted otherwise)

The General Radio 1657 Digibridge is a digital RLC meter, introduced in 1976 1.

Specifications

please add

Links

Internals

The 1657 applies a sine wave to the measured device and a precision resistor in series. The average voltage across each is measured using four different phases relative to the input sine wave, for a total of eight measurements. From these, the resistive and reactive components of the tested device are calculated and displayed.

The 1657 uses a 6503 CPU, which is a 28-pin version of the 6502 with the same instruction set, but fewer address lines. The 1657's firmware is contained in a 2Kx8 ROM, in later versions a 2716 EPROM. A 128-byte SRAM holds scratchpad and the stack.