The {{Title|General Radio 1822 Digital Voltmeter Calibrator}} was available in {{Catalog U}} only. | The {{Title|General Radio 1822 Digital Voltmeter Calibrator}} was available in {{Catalog U}} only. | ||
The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters. | The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters. | ||
as a precise voltage reference from 100 μV | It can be used as a precise laboratory voltage reference from 100 μV DC to 1111.1 V DC. | ||
The output voltage is displayed using a mechanical digital display with automatic units and decimal placement. It can set to any value with five-digit resolution, the resolution is adjustable. | |||
External programming is available through a rear-panel connector along with an internal automatic step function. | |||
Other features include an interference voltage function, polarity reversal, and adjustable output impedance. | |||
==Specifications== | ==Specifications== | ||
* '''Output Voltage:''' 100 μV | * '''Output Voltage:''' 100 μV DC to 1111.1 V DC in 7 decade ranges | ||
* '''Output Impedance:''' <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A | * '''Output Impedance:''' <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A 1 kΩ, 10 kΩ, or 100 kΩ output resistor can be added to above impedances by front-panel selector switch | ||
* '''Output Current:''' 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range | * '''Output Current:''' 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range | ||
* '''Accuracy:''' ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV) | * '''Accuracy:''' ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV) | ||