716-C

From GRWiki
Revision as of 19:29, 30 May 2024 by Peter (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search
General Radio 716-C
capacitance bridge
General Radio 716-C Capacitance Bridge

Available from 1948 to 1970

Manuals
Catalog History
 Document Year Page
Catalog L 1948 70
Catalog M 1951 69
Catalog N 1954 83
Catalog O 1956 18
Catalog P 1959 10
Catalog Q 1961 13
Catalog R 1963 38
Catalog S 1965 60
Catalog T 1968 82
Catalog U 1970 92
(All manuals in PDF format unless noted otherwise)

The General Radio 716-C Capacitance Bridge was introduced in Catalog L (1948) and remained available through Catalog U (1970).

The Type 716-C capacitance bridge uses the Schering bridge circuit and is direct reading at five frequencies (100 Hz, 1 kHz, 10 kHz, and 100 kHz). Dissipation factor is direct-reading at 1 kHz but the usable range is greater, see the manual.

Capacitance measurements can be made by the direct or substitution methods. When using the substitution method, a balancing capacitor is required that is 100 pF greater than the unknown, the Types 505 and 509 are recommended. Inductance and Q of large inductors can be measured by the substitution method.

The 716-C requires a generator and detector, the Types 1310-A or 1311-A oscillators and 1232-A null detector are recommended.

The Type 716-C is the last instrument in 716 series being replaced by the Type 1615-A.

Specifications

  • Capacitance Range: Direct method, 100 to 1150 pF; Substitution method 0.1 to 1050 pF
  • Accuracy: ±0.1%
  • Dissipation Factor Range: 0.00002 to 0.56
  • Accuracy: ±2%
  • Frequency: 30 Hz to 300 kHz; useful up to 1 MHz with corrections

Links

Photos