Type '''1900-A Wave Analyzer'''.
{{GR Product
|model=1900-A
|codes=1900-9801
|class=wave analyzer
|summary=Wave Analyzer
|image=GR 1900-A Close-Up.jpg
|caption=General Radio 1900-A Wave Analyzer
|series=1900
|introduced=1963
|discontinued=1973
|designers=
|manuals=
* [[Media:GR 1900-A Wave Analyzer 1900-0100-A 05_1965.pdf|General Radio 1900-A Wave Analyzer Manual 1900-0100-A 1965]]
{{Catalog History}}
}}
The {{Title|General Radio 1900-A Wave Analyzer}} was introduced in {{Catalog R}} and remained available through {{Catalog 1973}}.
 
The Type 1900-A Wave Analyzer is designed to measure and analyze electrical signals between 20 Hz and 54 KHz at bandwidths as narrow as 3 Hz.  When coupled with the recorder Type [[1521-B]] a permanent record of the spectrum of interest is produced. The 1900-A has a built in tracking generator whose frequency is synchronized with the analyzer frequency. This generator can be used as a source for bridges or networks, the output can be measured using its high selectivity to reduce noise and harmonics. Applications include analyzing filters and frequency response of amplifiers or networks.
The 1900-A is complex instrument but major components are housed in their respective shielded enclosures.  These subassemblies are connected using RCA and Jones type connectors simplifying service.  The 1900-A and the [[1521-B]] combination make up the Type [[1910-A]] Recording Wave Analyzer.


==Specifications==
==Specifications==
* '''Frequency Range:''' 20 to 54,000 Hz
* '''Accuracy of Calibration:''' ±(½% +5 Hz) up to 50 kHz;. ±1% beyond 50 kHz
* '''Incremental Frequency Dial (ΔF):''' ±100 Hz. accuracy is ±2 Hz below 2 kHz, ±5 Hz up to 50 kHz
* '''Selectivity:''' three bandwidths (3, 10, and 50 Hz) selected by switch
* '''Input Impedance:''' 1 MΩ on all ranges
* '''Input Voltage Range:''' 30 μV to 300 V full scale


==Links==
==Links==
* [[Media:GR Exp 1900-A 04_1964.pdf|Experimenter April 1964 describing Type 1900-A]]
* [[Media:GR Exp 1900-A 08_1966.pdf|Experimenter August 1966 describing Type 1900-A]]
* [[1910]]
* [[1910]]


==Photos==
==Photos==
<gallery>
<gallery>
GR-1910-ad-Electronics-1969-10-04.jpg
GR-1910-ad-Electronics-1969-10-04.jpg
GR 1900-A Front.jpg
GR 1900-A Case Side.jpg
GR 1900-A Case Rear.jpg
GR 1900-A Case Rear Close-Up.jpg
</gallery>
===Top View===
<gallery>
GR 1900-A 1 Inside Top.jpg
GR 1900-A 2 Inside Top Left.jpg
GR 1900-A 3 Inside Top Center.jpg
GR 1900-A 4 Inside Top Right.jpg
GR 1900-A 5 Inside Top Center Close-up.jpg
</gallery>
===Left View===
<gallery>
GR 1900-A 1 Inside Left Side.jpg
GR 1900-A 2 Inside Rear Board B.jpg
GR 1900-A 3 Inside Bottom Left  Board.jpg
</gallery>
===Right View===
<gallery>
GR 1900-A 1 Inside Right Side.jpg
GR 1900-A 2 Inside Right Top.jpg
GR 1900-A 3 Inside Right upper.jpg
GR 1900-A 4 Inside Right Lower.jpg
GR 1900-A 5 Inside Left Side Open.jpg
</gallery>
===Bottom View===
<gallery>
GR 1900-A 1 Inside Bottom.jpg
GR 1900-A 2 Inside Bottom Right.jpg
GR 1900-A 3 Inside Bottom Left Close-Up.jpg
GR 1900-A 4 Inside Bottom Left.jpg
GR 1900-A 5 Inside Bottom Left Open.jpg
GR 1900-A 6 Inside Bottom Open.jpg
GR 1900-A 7 Inside Bottom Left Open A.jpg
GR 1900-A 8 Inside Bottom Right Open.jpg
</gallery>
===Rear View===
<gallery>
GR 1900-A 1 Inside Rear Left.jpg
GR 1900-A 2 Inside Rear Right Upper.jpg
GR 1900-A 3 Inside Upper Left.jpg
GR 1900-A 4 Inside Rear Center.jpg
GR 1900-A 5 Inside Rear Upper Right.jpg
GR 1900-A 6 Inside Rear Lower Right.jpg
GR 1900-A 7 Inside Rear Right.jpg
</gallery>
</gallery>


[[Category:Wave Analyzer]]
[[Category:Wave Analyzer]]
__NOTOC__

Revision as of 19:03, 14 May 2024

General Radio 1900-A
Wave Analyzer
General Radio 1900-A Wave Analyzer

Available from 1963 to 1973

Manuals
Catalog History
 Document Year Page
Catalog R 1963 14
Catalog S 1965 28
Catalog T 1968 30
Catalog U 1970 44
Catalog73 1973 60
(All manuals in PDF format unless noted otherwise)

The General Radio 1900-A Wave Analyzer was introduced in Catalog R (1963) and remained available through Catalog 1973.

The Type 1900-A Wave Analyzer is designed to measure and analyze electrical signals between 20 Hz and 54 KHz at bandwidths as narrow as 3 Hz. When coupled with the recorder Type 1521-B a permanent record of the spectrum of interest is produced. The 1900-A has a built in tracking generator whose frequency is synchronized with the analyzer frequency. This generator can be used as a source for bridges or networks, the output can be measured using its high selectivity to reduce noise and harmonics. Applications include analyzing filters and frequency response of amplifiers or networks. The 1900-A is complex instrument but major components are housed in their respective shielded enclosures. These subassemblies are connected using RCA and Jones type connectors simplifying service. The 1900-A and the 1521-B combination make up the Type 1910-A Recording Wave Analyzer.

Specifications

  • Frequency Range: 20 to 54,000 Hz
  • Accuracy of Calibration: ±(½% +5 Hz) up to 50 kHz;. ±1% beyond 50 kHz
  • Incremental Frequency Dial (ΔF): ±100 Hz. accuracy is ±2 Hz below 2 kHz, ±5 Hz up to 50 kHz
  • Selectivity: three bandwidths (3, 10, and 50 Hz) selected by switch
  • Input Impedance: 1 MΩ on all ranges
  • Input Voltage Range: 30 μV to 300 V full scale