1602-B: Difference between revisions
(→Links) |
No edit summary |
||
Line 11: | Line 11: | ||
|manuals= | |manuals= | ||
* [[Media:GRwiki 1602-B Amitance Meter 1602-0110-J 1968.pdf|General Radio 1602-B Admittance Meter Manual 1602-0110-J 1968]] | * [[Media:GRwiki 1602-B Amitance Meter 1602-0110-J 1968.pdf|General Radio 1602-B Admittance Meter Manual 1602-0110-J 1968]] | ||
{{Catalog History}} | |||
* [[Media:GRwiki Catalog N OCR 1954.pdf|N (1954)]], p.102 | |||
* [[Media:GRwiki Catalog O OCR 1956.pdf|O (1956)]], p.40 | |||
* [[Media:GRwiki Catalog P OCR 1959.pdf|P (1959)]], p.34 | |||
* [[Media:GRwiki Catalog Q OCR 1961.pdf|Q (1961)]], p.34 | |||
* [[Media:GRwiki Catalog R OCR 1963.pdf|R (1963]], p.54 | |||
* [[Media:GRwiki Catalog S OCR 1965.pdf|S (1965)]], p.54 | |||
* [[Media:GRwiki Catalog T OCR 1968.pdf|T (1968)]], p.128 | |||
* [[Media:GRwiki Catalog U OCR 1970.pdf|U (1970)]], p.144 | |||
* [[Media:GRwiki Catalog 1973.pdf|Catalog73 (1973)]], p.210 | |||
* [[Media:GRwiki Catalog 1978.pdf|Catalog78 (1978)]], p.37 | |||
}} | }} | ||
The '''General Radio 1602-B UHF Admittance Meter''' was introduced in 1954 Catalog M and remained available through Catalog78 1978. | The '''General Radio 1602-B UHF Admittance Meter''' was introduced in 1954 Catalog M and remained available through Catalog78 1978. |
Revision as of 21:09, 29 March 2024
The General Radio 1602-B UHF Admittance Meter was introduced in 1954 Catalog M and remained available through Catalog78 1978.
The Type 1602-B is a null type instrument designed to measure admittance at UHF frequencies with coaxial systems. It achieves this by the use of three loops parallel connected to the detector. The three loops are coupled magnetically to the unknown admittance, conductance standard and susceptance standard lines which are driven directly by the generator. By adjusting these loops for a minimum (null) at the detector a direct reading from the three scales is determined. The 1602-B is an update from the 1602-A with an expanded "Multiplying Scale" and locking coaxial connectors. Supplied with the instrument are two multiplier plates that improve accuracy, resolution and range. The susceptance scale is direct reading only with matched serial numbered susceptance standards supplied with the instrument.
Specifications
RANGES OF MEASUREMENT
- Conductance: 0.01 to 4000 mS
- Susceptance: -4000 to +4000 mS
- Frequency: 40 to 1500 MHz, direct reading.
- Accuracy (conductance and susceptance): Up to 1000 MHz, from 0 to 20 mS, + (3% ± 0.2 mS) from 20 to 00 mS, +(3 √M% + 0.2 mS) where M is the scale-multiplying factor.
Links
- Experimenter describing Type 1602-B May 1960
- Experimenter describing Type 1602-B August 1953
- Experimenter describing Type 1602-A May 1950
Catalog History
Catalog | Year | Page Number |
---|---|---|
N | 1954 | 102 |
O | 1956 | 40 |
P | 1959 | 34 |
Q | 1961 | 34 |
R | 1963 | 54 |
S | 1965 | 54 |
T | 1968 | 128 |
U | 1970 | 144 |
73 | 1973 | 210 |
78 | 1978 | 37 |
Photos
-
-
-
-
-
-
-
-
-
-
-
Supplied Accessories
-
Multiplier Plates