1822: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
 
Line 7: Line 7:
|caption=General Radio 1822 Digital Voltmeter Calibrator
|caption=General Radio 1822 Digital Voltmeter Calibrator
|series=
|series=
|introduced=1970
|introduced=1969
|discontinued=1970
|discontinued=1970
|designers=
|designers=
Line 14: Line 14:
{{Catalog History}}
{{Catalog History}}
}}
}}
The {{Title|General Radio 1822 Digital Voltmeter Calibrator}} was available in {{Catalog U}} only.  
The {{Title|General Radio 1822 Digital Voltmeter Calibrator}} was introduced in {{Catalog T Supplement}} and available until {{Catalog U}}.  


The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters.
The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters.

Latest revision as of 09:20, 3 December 2024

General Radio 1822
Digital Voltmeter Calibrator
General Radio 1822 Digital Voltmeter Calibrator

Available from 1969 to 1970

Manuals
  • Manual Needed.
Catalog History
 Document Year Page
Catalog U 1970 257
(All manuals in PDF format unless noted otherwise)

The General Radio 1822 Digital Voltmeter Calibrator was introduced in Catalog T Supplement (1969) and available until Catalog U (1970).

The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters. It can be used as a precise laboratory voltage reference from 100 μV DC to 1111.1 V DC. The output voltage is displayed using a mechanical digital display with automatic units and decimal placement. It can set to any value with five-digit resolution, the resolution is adjustable.

External programming is available through a rear-panel connector along with an internal automatic step function. Other features include an interference voltage function, polarity reversal, and adjustable output impedance.

Specifications

  • Output Voltage: 100 μV DC to 1111.1 V DC in 7 decade ranges
  • Output Impedance: <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A 1 kΩ, 10 kΩ, or 100 kΩ output resistor can be added to above impedances by front-panel selector switch
  • Output Current: 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range
  • Accuracy: ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV)