General Radio 736-A
Wave Analyzer
General Radio 736-A Wave Analyzer

Available from 1939 to 1961

Manuals
Catalog History
 Document Year Page
Catalog K1 1939 166
Catalog K2 1942 118
Catalog K3 1944 102
Catalog K4 1945 102
Catalog L 1948 130
Catalog M 1951 162
Catalog N 1954 176
Catalog O 1956 224
Catalog P 1959 223
Catalog Q 1961 225
(All manuals in PDF format unless noted otherwise)

The General Radio 736-A Wave Analyzer was introduced in Catalog K1 (1939) and remained available through Catalog Q (1961).

The Type 736-A Wave Analyzer is used to measure the amplitude and frequency of components of a complex electrical waveform with 4 Hz bandwith. It performs this by the heterodyne method.

The band of frequencies between 20 Hz and 16 kHz are heterodyned or mixed with the local oscillator producing an intermediate frequency of 50 kHZ along with its sidebands. The frequencies applied to the input of the instrument are rejected by the 50 kHz tuned IF. A ballanced mixer cancels the local oscillator and the three stage crystal filter rejects the lower sideband. The upper sideband is then amplified by the following three stages and applied to the panel meter. The frequency dial is calibrated in the frequency applied to the panel meter. Under a cover at the base of the instrument are the calibration controls.

The Type 736-A can be used to measure harmonic distortion in amplifiers and radio receivers. It also finds use as null detector at audio frequencies, the measurement of hum and as a sharply tuned voltmeter.



Specifications

  • Frequency Range: 20 Hz to 16,000 kHz
  • Selectivity: 4 Hz
  • Voltage Range: 300 μV to 300 V
  • Volage Accuracy: ±5%
  • Input Impedance: 1 MΩ for direct voltage measurements; 100 kΩ when using input potentiometer