Line 1: Line 1:
Type '''1409-X 0.5uf Standard Capacitor'''.
{{GR Product
|model=1409-X
|codes=1409-1409-9721
|class=Capacitance Standards
|summary=0.5μF,0.05% Standard Capacitor
|image=Photo Needed.jpg
|caption=General Radio 1409-X Standard Capacitor
|series=1409
|introduced=1956
|discontinued=1970
|designers=
|manuals=
* '''Manual Needed.'''
{{Catalog History}}
}}
The {{Title|General Radio General Radio 1409-X Standard Capacitor}} was introduced in {{Catalog O}} and remained available through {{Catalog U}}.
 
The 1409-X is a fixed mica capacitor of very high stability for use as two or three-terminal capacitance standards in the laboratory. The capacitor is installed in an aluminum case along with Silica gel desiccant then sealed with hi-temp potting wax. A well located on the right side of the case for the insertion of a dial-type thermometer. Three jack-top binding posts are provided on the top of the case and removable plugs on the bottom, for convenient parallel connection without error. Each capacitor is supplied with a certificate of calibration giving two and three terminal measured capacitance. The Type 1409-X capacitor is mounted in the small case.
 
See [[1409]] to view all the capacitors in the series.


==Specifications==
==Specifications==
* '''Capacitance:''' 0.5 μF
* '''Series Inductance:''' 0.055 μH
* '''Adjustment Accuracy:''' within ±0.05%
* '''Stability:''' Capacitance change is less than 0.01% per year
* '''Temperature Coefficient:''' capacitance; +35 ± 10 ppm per degree between 10 °C and 70 °C
* '''Dissipation Factor:''' Less than 0.0003 at 1 kHz and 23 °C (see curves). Measured dissipation factor at 1 kHz is stated in the certificate to an accuracy of ±0.00005
* '''Leakage Resistance:''' 5000 ohm-farads or 100 GΩ, whichever is the lesser
* '''Maximum Voltage:''' 500 V peak below 35 kHz


==Links==
==Links==


* [[Media:GR Exp 1409 07_1957.pdf|Experimenter July 1957 describing Type 1409 Series]]


==Photos==
==Photos==
<gallery>
<gallery>
GR 1209 Figure 1.jpg|Change in capacitance as a function of frequency for typical Type 1409 capacitors.
GR 1209 Figure 2.jpg|Dissipation factor as a function of frequency.
GR 1409 Technical Drawing.jpg|Technical Drawing.
Photo Needed.jpg
GR 1409 Small Case Left.jpg| Thermometer Access
GR 1409 Small Case Right.jpg
GR 1409 Small Case Top.jpg
GR 1409 Small Case Bottom.jpg
GR 1409 Small Case Back.jpg
</gallery>
</gallery>


[[Category:Capacitance Standards]]
[[Category:Capacitance Standards]]

Revision as of 21:15, 21 June 2024

General Radio 1409-X
0.5μF,0.05% Standard Capacitor
General Radio 1409-X Standard Capacitor

Available from 1956 to 1970

Manuals
  • Manual Needed.
Catalog History
 Document Year Page
Catalog O 1956 156
Catalog P 1959 169
Catalog Q 1961 163
Catalog R 1963 152
Catalog S 1965 192
Catalog T 1968 89
Catalog U 1970 115
(All manuals in PDF format unless noted otherwise)

The General Radio General Radio 1409-X Standard Capacitor was introduced in Catalog O (1956) and remained available through Catalog U (1970).

The 1409-X is a fixed mica capacitor of very high stability for use as two or three-terminal capacitance standards in the laboratory. The capacitor is installed in an aluminum case along with Silica gel desiccant then sealed with hi-temp potting wax. A well located on the right side of the case for the insertion of a dial-type thermometer. Three jack-top binding posts are provided on the top of the case and removable plugs on the bottom, for convenient parallel connection without error. Each capacitor is supplied with a certificate of calibration giving two and three terminal measured capacitance. The Type 1409-X capacitor is mounted in the small case.

See 1409 to view all the capacitors in the series.

Specifications

  • Capacitance: 0.5 μF
  • Series Inductance: 0.055 μH
  • Adjustment Accuracy: within ±0.05%
  • Stability: Capacitance change is less than 0.01% per year
  • Temperature Coefficient: capacitance; +35 ± 10 ppm per degree between 10 °C and 70 °C
  • Dissipation Factor: Less than 0.0003 at 1 kHz and 23 °C (see curves). Measured dissipation factor at 1 kHz is stated in the certificate to an accuracy of ±0.00005
  • Leakage Resistance: 5000 ohm-farads or 100 GΩ, whichever is the lesser
  • Maximum Voltage: 500 V peak below 35 kHz