The '''General Radio 1611-B Capacitance Test Bridge''' was introduced in {{Catalog P}} and remained available through {{Catalog S}}. | The '''General Radio 1611-B Capacitance Test Bridge''' was introduced in {{Catalog P}} and remained available through {{Catalog S}}. | ||
The Type 1611-B self contained capacitance bridge operating at 60 Hz. It can be used at 120 Hz with an external oscillator and filter, the Type [[1214-D]] is recommended. Capacitance measurements at 120 cps are used primarily for the measurement of polarized electrolytic capacitors. A 6E5 "Eye" tube is used as the null indicator along with detector | The Type 1611-B self contained capacitance bridge operating at 60 Hz. It can be used at 120 Hz with an external oscillator and filter, the Type [[1214-D]] is recommended. Capacitance measurements at 120 cps are used primarily for the measurement of polarized electrolytic capacitors. A 6E5 "Eye" tube is used as the null indicator along with a detector sensitivity control. The bridge is suitable for testing of paper, mica, and electrolytic capacitors with 1% accuracy. | ||
==Specifications== | ==Specifications== |
Revision as of 22:43, 18 April 2024
The General Radio 1611-B Capacitance Test Bridge was introduced in Catalog P (1959) and remained available through Catalog S (1965).
The Type 1611-B self contained capacitance bridge operating at 60 Hz. It can be used at 120 Hz with an external oscillator and filter, the Type 1214-D is recommended. Capacitance measurements at 120 cps are used primarily for the measurement of polarized electrolytic capacitors. A 6E5 "Eye" tube is used as the null indicator along with a detector sensitivity control. The bridge is suitable for testing of paper, mica, and electrolytic capacitors with 1% accuracy.
Specifications
- Capacitance Range: 0 to 11,000 μF in eight ranges
- Test-Voltage Frequency: 60 Hz with internal supply and filter; 120 cps with external supply on four higher multipliers
- Capacitance Accuracy: ±(1% + l μμF) over the entire range
- Dissipation-Factor Range: 0 to 60% at 60 cps (x f/60 at other frequencies)
- Dissipation-Factor Accuracy: ±(2% of dial reading + 0.05% x f/60 dissipation factor)