1605-A: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 11: Line 11:
|manuals=
|manuals=
* [[Media:GRwiki 1605-A Impedance Comparator 1605-0100-E 04-1961.pdf|General Radio 1605-A Impedance Comparator Manual 1605-01100-E 1961]]
* [[Media:GRwiki 1605-A Impedance Comparator 1605-0100-E 04-1961.pdf|General Radio 1605-A Impedance Comparator Manual 1605-01100-E 1961]]
{{Catalog History}}
* [[Media:GRwiki Catalog O OCR 1956.pdf|O (1956)]], p.32
* [[Media:GRwiki Catalog P OCR 1959.pdf|P (1959)]], p.26
* [[Media:GRwiki Catalog Q OCR 1961.pdf|Q (1961)]], p.26
* [[Media:GRwiki Catalog R OCR 1963.pdf|R (1963]], p.50
* [[Media:GRwiki Catalog S OCR 1965.pdf|S (1965)]], p.46
* [[Media:GRwiki Catalog T OCR 1968.pdf|T (1968)]], p.70
}}
}}
The '''General Radio 1605-A Impedance Comparator''' was introduced in 1956 Catalog O and remained available through Catalog T 1968.
The '''General Radio 1605-A Impedance Comparator''' was introduced in 1956 Catalog O and remained available through Catalog T 1968.



Revision as of 22:21, 31 March 2024

General Radio 1605-A
impedance comparator
General Radio 1605-A Impedance Comparator

Available from 1956 to 1968

Manuals
Catalog History
 Document Year Page
Catalog O 1956 32
Catalog P 1959 26
Catalog Q 1961 26
Catalog R 1963 50
Catalog S 1965 46
Catalog T 1968 70
(All manuals in PDF format unless noted otherwise)


The General Radio 1605-A Impedance Comparator was introduced in 1956 Catalog O and remained available through Catalog T 1968.

The Type 1605-A can be used to compare pairs of components for magnitude and phase difference. It displays the results on separate meters (magnitude, phase). Using the built-in guard circuit during measurement, the unknown can be remotely located during measurement. Some other applications include checking ganged potentiometers or variable capacitors for tracking, checking balances of transformer windings, or adjusting coils to a standard.

Specifications

  • Impedance Ranges:
    • Resistance: 2 Ω to 20 MΩ
    • Capacitance: 40 pF to 500 μF
    • Inductance: 20 μh to 10,000 H
  • Internal Oscillator Frequencies: 100 Hz, 1 kHz, 10 kHz, 100 kHz
  • Meter Ranges:
    • Impedance Magnitude Difference: ±0.3%, ± 1%, ±3%, ± 10% full scale
    • Phase Angle Difference: ±0.003, ±0.01, ±0.03, ± 0.1 radian full scale
  • Voltage Across Standard and Unknown: approx. 0.15 V