724-A: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
 
(One intermediate revision by one other user not shown)
Line 16: Line 16:
The {{Title|General Radio 724-A Precision Wavemeter}} was introduced in {{Catalog J}} and remained available through {{Catalog K4}}.  
The {{Title|General Radio 724-A Precision Wavemeter}} was introduced in {{Catalog J}} and remained available through {{Catalog K4}}.  


The Type 724-A is an absorption type wavemeter whose range is from 17 kHz to 50 MHz using seven coils with an accuracy of ±0.25%. The primary application of the 724-A is oscillator and transmitter frequency measurement.  
The Type 724-A is an absorption-type wavemeter with a range from 17 kHz to 50 MHz using seven coils with an accuracy of ±0.25%. The primary application of the 724-A is oscillator and transmitter frequency measurement.  


The capacitor used is similar to the Type [[722-N]] Precision Capacitor. Its connected to the inductor through a [[774|Type 774 coaxial connector]]. Resonance is achieved using one of three methods, absorption method or reaction method and beat method.  Using the panel galvanometer the the absorption method is employed. Regardless of the method, once resonance is achieved its a matter of using the correct chart for the coil in use to determine the frequency.  
The capacitor used is similar to the [[722-N|Type 722-N Precision Capacitor]]. It is connected to the inductor through a [[774|Type 774 coaxial connector]].  
Resonance is achieved using one of three methods – absorption, reaction, or beat method.   
Using the panel galvanometer, the absorption method is employed.  
Regardless of the method, once resonance is achieved, it is a matter of using the correct chart for the coil in use to determine the frequency.  


The Type 724-A is supplied as a collection of calibrated components sharing the same serial number. That is the capacitor, all the inductors, and calibration charts must have the same serial number to achieve the stated accuracy. The 724-A replaced the [[224-A]] Wavemeter with improved coils and a galvanometer replacing the thermocouple type indicator. The wavemeter is furnished with a wooden storage case, fitted with lock and carrying handle. This case has compartments for holding the capacitor, inductors, and calibration charts. The 724-A uses a 1G4GT/G Triode vacuum tube, resulting in the need for a 1.5 V battery for the heater.
Type 724-A was supplied as a collection of calibrated components sharing the same serial number, i.e. the capacitor, inductors, and calibration charts must have the same serial number to achieve the stated accuracy.  
 
The 724-A replaced the [[224-A|224-A Wavemeter]] with improved coils and a galvanometer replacing the thermocouple type indicator.  
The wavemeter was furnished with a wooden storage case, fitted with lock and carrying handle.  
This case has compartments for holding the capacitor, inductors, and calibration charts.  
The 724-A uses a 1G4GT/G triode vacuum tube, resulting in the need for a 1.5 V battery for the heater.
 
Note: The instrument shown in the photos below uses a 1LE3 loctal battery triode. It appears to be a production change.


==Specifications==
==Specifications==

Latest revision as of 18:17, 30 May 2024

General Radio 724-A
Precision Wavemeter
General Radio 724-B Precision Wavemeter

Available from 1936 to 1945

Manuals
Catalog History
 Document Year Page
Catalog J 1936 57
Catalog K1 1939 161
Catalog K2 1942 176
Catalog K3 1944 147
Catalog K4 1945 149
(All manuals in PDF format unless noted otherwise)

The General Radio 724-A Precision Wavemeter was introduced in Catalog J (1936) and remained available through Catalog K4 (1945).

The Type 724-A is an absorption-type wavemeter with a range from 17 kHz to 50 MHz using seven coils with an accuracy of ±0.25%. The primary application of the 724-A is oscillator and transmitter frequency measurement.

The capacitor used is similar to the Type 722-N Precision Capacitor. It is connected to the inductor through a Type 774 coaxial connector. Resonance is achieved using one of three methods – absorption, reaction, or beat method. Using the panel galvanometer, the absorption method is employed. Regardless of the method, once resonance is achieved, it is a matter of using the correct chart for the coil in use to determine the frequency.

Type 724-A was supplied as a collection of calibrated components sharing the same serial number, i.e. the capacitor, inductors, and calibration charts must have the same serial number to achieve the stated accuracy.

The 724-A replaced the 224-A Wavemeter with improved coils and a galvanometer replacing the thermocouple type indicator. The wavemeter was furnished with a wooden storage case, fitted with lock and carrying handle. This case has compartments for holding the capacitor, inductors, and calibration charts. The 724-A uses a 1G4GT/G triode vacuum tube, resulting in the need for a 1.5 V battery for the heater.

Note: The instrument shown in the photos below uses a 1LE3 loctal battery triode. It appears to be a production change.

Specifications

  • Frequency Range: 16 kHz to 50 MHz using seven inductors
  • Accuracy: ±0.25% between 50 kHz and 50 MHz, 1% between 16 and 50 kHz
  • Inductors: 7 coils wound on isolantite forms and enclosed in molded bakelite cases
  • Capacitor: Precision worm-drive type similar to Type 722
  • Resonance Indicator: galvanometer and vacuum tube rectifier combination

Links

Photos