716-C: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
 
(5 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{GR Product
{{GR Product
|model=716-C
|model=716-C
|codes=BOSOM,BONUS
|class=impedance measurement
|class=impedance measurement
|summary=capacitance bridge
|summary=capacitance bridge
Line 11: Line 12:
|manuals=
|manuals=
* [[Media:GR 716C Capacitance Bridge 681-K 11-1959.pdf|General Radio 716-C Capacitance Bridge Manual 681-K 1959]]
* [[Media:GR 716C Capacitance Bridge 681-K 11-1959.pdf|General Radio 716-C Capacitance Bridge Manual 681-K 1959]]
* [[Media:GR 716-C Capacitance Bridge 0716-0140-L 05-1963.pdf|General Radio 716-C Capacitance Bridge Manual 0716-0140-L 1963]]
{{Catalog History}}
}}
}}
The {{Title|General Radio 716-C Capacitance Bridge}} was introduced in {{Catalog L}} and remained available through {{Catalog U}}.  
The {{Title|General Radio 716-C Capacitance Bridge}} was introduced in {{Catalog L}} and remained available through {{Catalog U}}.  


The Type 716-C capacitance bridge uses the Schering circuit and is direct reading at five frequencies, 100 Hz, 1 kHz, 10 kHz, and 100 kHz. Dissipation factor is direct reading at 1 kHz but the usable range is greater, see the manual. Capacitance measurements can be made by the direct or substitution methods. When using the substitution method a balancing capacitor is required that is 100 pF greater than the unknown, the Types 505 and 509 are recommended. Inductance and Q of large inductors can be measured by the substitution method. The 716-C requires a generator and detector, the Types [[1310-A]] or [[1311-A]] oscillators and [[1232-A]] null detector are recommended.
The Type 716-C capacitance bridge uses the [[wikipedia:Schering bridge|Schering bridge circuit]] and is direct reading at five frequencies (100 Hz, 1 kHz, 10 kHz, and 100 kHz). Dissipation factor is direct-reading at 1 kHz but the usable range is greater, see the manual.  
 
Capacitance measurements can be made by the direct or substitution methods. When using the substitution method, a balancing capacitor is required that is 100 pF greater than the unknown, the Types [[505]] and [[509]] are recommended. Inductance and Q of large inductors can be measured by the substitution method.  
 
The 716-C requires a generator and detector, the Types [[1310-A]] or [[1311-A]] oscillators and [[1232-A]] null detector are recommended.  
 
The Type 716-C is the last instrument in 716 series being replaced by the Type [[1615-A]].


==Specifications==
==Specifications==
Line 26: Line 35:
==Links==
==Links==


* [[Media:GR Exp 716-A 12_1936.pdf|Experimenter describing Type 716-A December 1936]]
* [[Media:GR Exp 716-A 12_1936.pdf|Experimenter December 1936 describing Type 716-A]]
* [[Media:GR Exp 716-B 04_1942.pdf|Experimenter describing Type 716-B April 1942]]
* [[Media:GR Exp 716-B 04_1942.pdf|Experimenter April 1942 describing Type 716-B]]
* [[Media:GR Exp 716-C 04_1947.pdf|Experimenter describing Type 716-C April 1947]]
* [[Media:GR Exp 716-C 04_1947.pdf|Experimenter April 1947 describing Type 716-C]]


==Photos==
==Photos==

Latest revision as of 19:29, 30 May 2024

General Radio 716-C
capacitance bridge
General Radio 716-C Capacitance Bridge

Available from 1948 to 1970

Manuals
Catalog History
 Document Year Page
Catalog L 1948 70
Catalog M 1951 69
Catalog N 1954 83
Catalog O 1956 18
Catalog P 1959 10
Catalog Q 1961 13
Catalog R 1963 38
Catalog S 1965 60
Catalog T 1968 82
Catalog U 1970 92
(All manuals in PDF format unless noted otherwise)

The General Radio 716-C Capacitance Bridge was introduced in Catalog L (1948) and remained available through Catalog U (1970).

The Type 716-C capacitance bridge uses the Schering bridge circuit and is direct reading at five frequencies (100 Hz, 1 kHz, 10 kHz, and 100 kHz). Dissipation factor is direct-reading at 1 kHz but the usable range is greater, see the manual.

Capacitance measurements can be made by the direct or substitution methods. When using the substitution method, a balancing capacitor is required that is 100 pF greater than the unknown, the Types 505 and 509 are recommended. Inductance and Q of large inductors can be measured by the substitution method.

The 716-C requires a generator and detector, the Types 1310-A or 1311-A oscillators and 1232-A null detector are recommended.

The Type 716-C is the last instrument in 716 series being replaced by the Type 1615-A.

Specifications

  • Capacitance Range: Direct method, 100 to 1150 pF; Substitution method 0.1 to 1050 pF
  • Accuracy: ±0.1%
  • Dissipation Factor Range: 0.00002 to 0.56
  • Accuracy: ±2%
  • Frequency: 30 Hz to 300 kHz; useful up to 1 MHz with corrections

Links

Photos