1521-B: Difference between revisions

From GRWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(5 intermediate revisions by 2 users not shown)
Line 1: Line 1:
Type '''1521-B Graphic Level Recorder'''.
{{GR Product
|model=1521-B
|codes=1521-9802
|class=recorders and dial Drives
|summary=Graphic Level Recorder
|image=GR 1521-B Close-Up.jpg
|caption=General Radio 1521-B Graphic Level Recorder
|series=1521
|introduced=1965
|discontinued=1978+
|designers=
|manuals=
* [[Media:GR 1521-B Graphic Level Recorder 1521-0150-D 03_1969.pdf|General Radio 1521-B Graphic Level Recorder Manual 1521-0150-D 1969]]
{{Catalog History}}
}}
The {{Title|General Radio 1521-B Graphic Level Recorder}} was introduced in {{Catalog S}} and remained available through {{Catalog 1978}}.
 
The Type 1521-B Graphic Level Recorder provides a permanent record of AC voltage versus time. With the addition of the [[1521-P10]] Drive Unit,  the chart paper feed is now synchronized with the paired instrument.  Is essentially converts the recorder from AC voltage versus time to AC voltage versus frequency.  Some examples of this are the Type [[1911]] Recording Sound and Vibration System and the Type [[1640-A]] Slotted Line Recorder system, and [[1910]] Recording Wave Analyzer.
 
The response to the input AC voltage is logarithmic so its linear in decibels. Using the supplied 40 dB potentiometer and four inch chart paper, travel of the pen is 10 dB per inch.  Other potentiometers are available, the Type [[1521-P1]] 20 dB and the Type [[1521-P3]] 80 dB. A special potentiometer for DC recording is available as the Type [[1521-P4]]. These potentiometers are part of the feedback circuit that position the pen, the pen being the wiper for the potentiometer. The built in attenuator adds additional
attenuation to the potentiometer.
 
A list of accessories for the 1521-A include high and low speed motors, custom chart paper, and drive links for specific GR instruments.
 


==Specifications==
==Specifications==
* '''Recording Range:''' 40 dB full scale with standard potentiometer
* '''Frequency Range:''' 7 Hz to 200 kHz
* '''Maximum Input Voltage:''' 100 V AC
* '''Input Attenuator:''' 60 dB in 10 dB steps
* '''Input Impedance:''' 10,000 Ω for AC level recording; 1000 Ω for DC


==Links==
==Links==


* [[Media:GR Exp 1521-A 06_1959.pdf|Experimenter June 1959 describing Type 1521-A]]


==Photos==
==Photos==
<gallery>
<gallery>
GR 1521-B Front.jpg
GR 1521-B Left Side.jpg
GR 1521-B Rear.jpg
GR 1521-B Inside Left.jpg
GR 1521-B Inside Rear A.jpg
GR 1521-B Inside Top.jpg
GR 1521-B Inside Board.jpg
GR 1521-B Inside Bottom A.jpg
GR 1521-B Pen Motor.jpg
GR 1521-B Power Supply Top.jpg
GR 1521-B Power Supply Bottom.jpg
GR 1521-B Inside Transmission.jpg
gr 1521-B P10 Drive Unit.jpg
GR-1910-ad-Electronics-1969-10-04.jpg
GR-1910-ad-Electronics-1969-10-04.jpg
</gallery>
</gallery>


[[Category:Recorders and Dial Drives]]
[[Category:Recorders and Dial Drives]]

Latest revision as of 21:11, 17 June 2024

General Radio 1521-B
Graphic Level Recorder
General Radio 1521-B Graphic Level Recorder

Available from 1965 to 1978+

Manuals
Catalog History
 Document Year Page
Catalog S 1965 178
Catalog T 1968 40
Catalog U 1970 51
Catalog73 1973 49
Catalog78 1978 173
(All manuals in PDF format unless noted otherwise)

The General Radio 1521-B Graphic Level Recorder was introduced in Catalog S (1965) and remained available through Catalog 1978.

The Type 1521-B Graphic Level Recorder provides a permanent record of AC voltage versus time. With the addition of the 1521-P10 Drive Unit, the chart paper feed is now synchronized with the paired instrument. Is essentially converts the recorder from AC voltage versus time to AC voltage versus frequency. Some examples of this are the Type 1911 Recording Sound and Vibration System and the Type 1640-A Slotted Line Recorder system, and 1910 Recording Wave Analyzer.

The response to the input AC voltage is logarithmic so its linear in decibels. Using the supplied 40 dB potentiometer and four inch chart paper, travel of the pen is 10 dB per inch. Other potentiometers are available, the Type 1521-P1 20 dB and the Type 1521-P3 80 dB. A special potentiometer for DC recording is available as the Type 1521-P4. These potentiometers are part of the feedback circuit that position the pen, the pen being the wiper for the potentiometer. The built in attenuator adds additional attenuation to the potentiometer.

A list of accessories for the 1521-A include high and low speed motors, custom chart paper, and drive links for specific GR instruments.


Specifications

  • Recording Range: 40 dB full scale with standard potentiometer
  • Frequency Range: 7 Hz to 200 kHz
  • Maximum Input Voltage: 100 V AC
  • Input Attenuator: 60 dB in 10 dB steps
  • Input Impedance: 10,000 Ω for AC level recording; 1000 Ω for DC

Links

Photos