1409-Y: Difference between revisions
No edit summary |
(→Photos) |
||
(2 intermediate revisions by 2 users not shown) | |||
Line 21: | Line 21: | ||
A well located on the right side of the case allows insertion of a dial-type thermometer. | A well located on the right side of the case allows insertion of a dial-type thermometer. | ||
Three jack-top binding posts on the top of the case and removable plugs on the bottom are provided for convenient parallel connection without error. | Three jack-top binding posts on the top of the case and removable plugs on the bottom are provided for convenient parallel connection without error. | ||
Each capacitor is supplied with a certificate of calibration giving two- and three-terminal measured capacitance. | Each capacitor is supplied with a certificate of calibration giving two- and three-terminal measured capacitance. | ||
The Type 1409-Y capacitor is mounted in the large case. | The Type 1409-Y capacitor is mounted in the large case. | ||
Line 47: | Line 48: | ||
GR 1409 Technical Drawing.jpg|Technical Drawing. | GR 1409 Technical Drawing.jpg|Technical Drawing. | ||
GR 1409-Y Front.jpg | GR 1409-Y Front.jpg | ||
GR 1409-Y Front A.jpg | |||
GR 1409-Y Rear Large Case.jpg | |||
</gallery> | </gallery> | ||
[[Category:Capacitance Standards]] | [[Category:Capacitance Standards]] |
Latest revision as of 17:06, 12 July 2024
The General Radio 1409-Y Standard Capacitor was introduced in Catalog O (1956) and remained available through Catalog 1978.
The 1409-Y is a fixed mica capacitor of very high stability for use as two- or three-terminal capacitance standard in the laboratory. The capacitor is installed in an aluminum case along with silica-gel desiccant, then sealed with high-temperature potting wax.
A well located on the right side of the case allows insertion of a dial-type thermometer. Three jack-top binding posts on the top of the case and removable plugs on the bottom are provided for convenient parallel connection without error.
Each capacitor is supplied with a certificate of calibration giving two- and three-terminal measured capacitance. The Type 1409-Y capacitor is mounted in the large case.
See 1409 to view all the capacitors in the series.
Specifications
- Capacitance: 1.0 μF
- Series Inductance: 0.070 μH
- Adjustment Accuracy: within ±0.05%
- Stability: Capacitance change is less than 0.01% per year
- Temperature Coefficient: capacitance; +35 ± 10 ppm per degree between 10 °C and 70 °C
- Dissipation Factor: Less than 0.0003 at 1 kHz and 23 °C (see curves). Measured dissipation factor at 1 kHz is stated in the certificate to an accuracy of ±0.00005
- Leakage Resistance: 5000 ohm-farads or 100 GΩ, whichever is the lesser
- Maximum Voltage: 500 V peak below 17 kHz