The {{Title|General Radio 1454-AH Decade Voltage Divider}} was introduced in {{Catalog P}} and remained available through {{Catalog S}}. | The {{Title|General Radio 1454-AH Decade Voltage Divider}} was introduced in {{Catalog P}} and remained available through {{Catalog S}}. | ||
Type 1454-AH is a four-stage voltage divider using the [[wikipedia:Kelvin–Varley divider|Kelvin–Varley divider]] circuit. The use of four special [[510|Type 510 resistance units]] achieve this. It features a constant 100 kΩ input resistance, with a maximum input voltage of 700 V at audio frequencies. | |||
Applications for the 1454-AH include voltmeter calibration, measurements of gain and attenuation, and turns ratio of transformers. | Applications for the 1454-AH include voltmeter calibration, measurements of gain and attenuation, and turns ratio of transformers. | ||
Type 1454-AH is identical to the Type [[1454-A]] except that all resistance values are increased by a factor of 10. Ungrounded or grounded operation is possible with a link at the input terminals. | |||
==Specifications== | ==Specifications== | ||
* '''Voltage Ratio:''' .0001 to 1.0000 in | * '''Voltage Ratio:''' 0.0001 to 1.0000 in steps of 0.0001 | ||
* '''Accuracy: ±0.04% of indicated ratio | * '''Accuracy:''' ±0.04% of indicated ratio | ||
* '''Frequency Characteristics:''' frequency error is <0.1% at 2 kHz if the external capacitance placed across the output terminals is less than 50 pF | * '''Frequency Characteristics:''' frequency error is <0.1% at 2 kHz if the external capacitance placed across the output terminals is less than 50 pF | ||
* '''Input Resistance:''' 100 kΩ | * '''Input Resistance:''' 100 kΩ | ||
* '''Output Resistance:''' Varies with output setting | * '''Output Resistance:''' Varies with output setting | ||
* '''Maximum Input Voltage:''' | * '''Maximum Input Voltage:''' 700 V<sub>RMS</sub>, limited to 350 V<sub>RMS</sub> for maximum accuracy | ||
==Links== | ==Links== | ||
* [[Media:GR Exp 1454-A_AH 07_1959.pdf|Experimenter July 1959 describing Type]] | * [[Media:GR Exp 1454-A_AH 07_1959.pdf|Experimenter July 1959 describing Type 1454-AH]] | ||
==Photos== | ==Photos== | ||
Line 70: | Line 71: | ||
GR 1454-AH Inside Top Decade 4.jpg | GR 1454-AH Inside Top Decade 4.jpg | ||
GR 1454-AH Inside Rear Decade 4.jpg | GR 1454-AH Inside Rear Decade 4.jpg | ||
GR 1454-AH Inside | GR 1454-AH Inside Bottom Decade 4.jpg | ||
</gallery> | </gallery> | ||
Latest revision as of 01:56, 19 May 2024
The General Radio 1454-AH Decade Voltage Divider was introduced in Catalog P (1959) and remained available through Catalog S (1965).
Type 1454-AH is a four-stage voltage divider using the Kelvin–Varley divider circuit. The use of four special Type 510 resistance units achieve this. It features a constant 100 kΩ input resistance, with a maximum input voltage of 700 V at audio frequencies.
Applications for the 1454-AH include voltmeter calibration, measurements of gain and attenuation, and turns ratio of transformers.
Type 1454-AH is identical to the Type 1454-A except that all resistance values are increased by a factor of 10. Ungrounded or grounded operation is possible with a link at the input terminals.
Specifications
- Voltage Ratio: 0.0001 to 1.0000 in steps of 0.0001
- Accuracy: ±0.04% of indicated ratio
- Frequency Characteristics: frequency error is <0.1% at 2 kHz if the external capacitance placed across the output terminals is less than 50 pF
- Input Resistance: 100 kΩ
- Output Resistance: Varies with output setting
- Maximum Input Voltage: 700 VRMS, limited to 350 VRMS for maximum accuracy