1231-P5: Difference between revisions

From GRWiki
Jump to navigation Jump to search
(Created page with "Type 1231-P5 Category:Impedance measurement")
 
No edit summary
 
(3 intermediate revisions by one other user not shown)
Line 1: Line 1:
Type 1231-P5
{{GR Product
|model=1231-P5
|codes=ALDER,ADOBE
|class=detectors amplifiers
|summary=Adjustable Filter
|image=GR 1231-P5 Close-up.jpg
|caption=General Radio 1231-P5 Adjustable Filter
|series=1231
|introduced=1951
|discontinued=1959
|designers=
|manuals=
* '''Manual Needed'''
{{Catalog History}}
}}
The {{Title|General Radio 1231-P5 Adjustable Filter}} was introduced in {{Catalog M}} and remained available through {{Catalog P}}.


[[Category:Impedance measurement]]
The Type 1231-P5 is designed as an accessory filter for the Type [[1231-B]] Amplifier and Null Detector.
Its purpose is to increase selectivity, reduce noise and harmonics.
 
The internal filters are parallel-resonant circuits, consisting of toroidal inductors and fixed capacitors for the 11 frequencies.
Four additional switch positions are provided for selecting toroid inductors L1 thru L4 only.
A pair of terminals for an external capacitor can be used for tuning the filter to any frequency from 20 Hz to 100 kHz.
 
'''CAUTION:''' The Type 1231-P5 filter should not be subjected to any DC current.
 
==Specifications==
* '''Filter Frequencies:''' 11 settings; 50, 100, 200, 500 Hz; 1, 2, 5, 10, 20, 50, 100 kHz
* '''Frequency Calibration:''' Within ±2%
* '''Selectivity:''' Better than 30 dB against the second harmonic
 
==Links==
 
* [[Media:GR Exp 1231-P5 06_1951.pdf|Experimenter June 1951 describing Type 1231-P5]]
 
==Photos==
<gallery>
GR 1231-P5 Front.jpg
GR 1231-P5 Left.jpg
GR 1231-P5 Rear.jpg
GR 1231-P5 Right.jpg
GR 1231-P5 Top Close-up.jpg
GR 1231-P5 Bottom.jpg
GR 1231-P5 Bottom Close-up.jpg
</gallery>
 
[[Category:Detectors Amplifiers]]

Latest revision as of 22:43, 30 April 2024

General Radio 1231-P5
Adjustable Filter
General Radio 1231-P5 Adjustable Filter

Available from 1951 to 1959

Manuals
  • Manual Needed
Catalog History
 Document Year Page
Catalog M 1951 99
Catalog N 1954 108
Catalog O 1956 64
Catalog P 1959 62
(All manuals in PDF format unless noted otherwise)

The General Radio 1231-P5 Adjustable Filter was introduced in Catalog M (1951) and remained available through Catalog P (1959).

The Type 1231-P5 is designed as an accessory filter for the Type 1231-B Amplifier and Null Detector. Its purpose is to increase selectivity, reduce noise and harmonics.

The internal filters are parallel-resonant circuits, consisting of toroidal inductors and fixed capacitors for the 11 frequencies. Four additional switch positions are provided for selecting toroid inductors L1 thru L4 only. A pair of terminals for an external capacitor can be used for tuning the filter to any frequency from 20 Hz to 100 kHz.

CAUTION: The Type 1231-P5 filter should not be subjected to any DC current.

Specifications

  • Filter Frequencies: 11 settings; 50, 100, 200, 500 Hz; 1, 2, 5, 10, 20, 50, 100 kHz
  • Frequency Calibration: Within ±2%
  • Selectivity: Better than 30 dB against the second harmonic

Links

Photos