* '''RANGES OF MEASUREMENT'''
* '''RANGES OF MEASUREMENT'''
* '''Voltage:''' ±30, 100, and 300 mV; ±1, 3, and 10 V; dc full scale.
* '''Voltage:''' ±30, 100, and 300 mV; ±1, 3, and 10 V; dc full scale.
* '''Current:''' ±1 mA dc, full scale, to ±300 pA full scale; Measured in terms of voltage.
* '''Current:''' ±1 mA dc, full scale, to ±0.3 pA full scale; Measured in terms of voltage.
* '''Resistance:''' direct reading from 300  kΩ to 10 TΩ full scale; There are 16 ranges, two per decade. Voltage across the unknown resistance is 9.1 V.
* '''Resistance:''' direct reading from 300  kΩ to 10 TΩ full scale; There are 16 ranges, two per decade. Voltage across the unknown resistance is 9.1 V.
* '''ACCURACY'''
* '''ACCURACY'''
* '''Voltage:''' ±2% of full scale on the five highest ranges, ±4% of full scale on the 30 mV range.
* '''Voltage:''' ±2% of full scale on the five highest ranges, ±4% of full scale on the 30 mV range.
* '''Current:''' ±3% of full scale from 1 mA to 1 nA, ±10% of full scale from 300 pA to 0.3 pA.
* '''Current:''' ±3% of full scale from 1 mA to 1 nA, ±10% of full scale from 300 pA to 0.3 pA.
* '''Resistance:''' ±3% from 300 kΩ to 10 MΩ at full scale (low-resistance end), ±8% from 3 MΩ to 10 TΩ.