The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters. It can be used
The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters.
as a precise voltage reference from 100 μV dc to 1111.1 V dc for laboratory use. The output voltage can set to any value with five digit resolution, the resolution is adjustable. The output voltage is displayed using a mechanical digital display with automatic units and decimal placement. External programming is available through a rear panel connector along with an internal automatic step function. Other features include an interference voltage function, polarity reversal, and adjustable output impedance.  
It can be used as a precise laboratory voltage reference from 100 μV DC to 1111.1 V DC.  
The output voltage is displayed using a mechanical digital display with automatic units and decimal placement. It can set to any value with five-digit resolution, the resolution is adjustable.  
 
External programming is available through a rear-panel connector along with an internal automatic step function.  
Other features include an interference voltage function, polarity reversal, and adjustable output impedance.  


==Specifications==
==Specifications==


* '''Output Voltage:''' 100 μV dc to 1111.1 V dc in 7 decade ranges
* '''Output Voltage:''' 100 μV DC to 1111.1 V DC in 7 decade ranges
* '''Output Impedance:''' <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A l kΩ, 10 kΩ, or 100 kΩ output resistor can be added to above impedances by front-panel selector switch
* '''Output Impedance:''' <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A 1 kΩ, 10 kΩ, or 100 kΩ output resistor can be added to above impedances by front-panel selector switch
* '''Output Current: 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range
* '''Output Current:''' 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range
* '''Accuracy:''' ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV)
* '''Accuracy:''' ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV)


Line 34: Line 38:
<gallery>
<gallery>
GR 1822 Front.jpg
GR 1822 Front.jpg
GR 1822 DVM calibrator.jpg
GR 1822 Side.jpg
GR 1822 Side.jpg
GR 1822 Rear.jpg
GR 1822 Rear.jpg
Line 45: Line 50:
GR 1822 Inside Top Rear Panel.jpg
GR 1822 Inside Top Rear Panel.jpg
GR 1822 Inside Bottom.jpg
GR 1822 Inside Bottom.jpg
GR 1822 Inside Bottom Card Connectors.jpg
GR 1822 Inside Bottom Front Left.jpg
GR 1822 Inside Bottom Front Left.jpg
GR 1822 Inside Bottom Front Panel.jpg
GR 1822 Inside Bottom Front Panel.jpg
GR 1822 Inside Bottom Middle Left.jpg
GR 1822 Inside Bottom Middle Left.jpg
</gallery>
</gallery>


Line 55: Line 58:


<gallery>
<gallery>
GR 1822-2720 Component.jpg
GR 1822-2720 Component.jpg|1822-2720
GR 1822-2720 Foil.jpg
GR 1822-2720 Foil.jpg|1822-2720
GR 1822-2731 Component.jpg
GR 1822-2731 Component.jpg|1822-2731
GR 1822-2731 Foil.jpg
GR 1822-2731 Foil.jpg|1822-2731
GR 1822-4750 Component.jpg
GR 1822-4750 Component.jpg|1822-4750
GR 1822-4750 Foil.jpg
GR 1822-4750 Foil.jpg|1822-4750
GR 1822-4756 Component.jpg
GR 1822-4756 Component.jpg|1822-4756
GR 1822-4756 Foil.jpg
GR 1822-4756 Foil.jpg|1822-4756
GR 1822-4761 Component.jpg
GR 1822-4761 Component.jpg|1822-4761
GR 1822-4761 Foil.jpg
GR 1822-4761 Foil.jpg|1822-4761
GR 1822-4766 Component.jpg
GR 1822-4766 Component.jpg|1822-4766
GR 1822-4766 Foil.jpg
GR 1822-4766 Foil.jpg|1822-4766
GR 1822-4770 Component.jpg
GR 1822-4770 Component.jpg|1822-4770
GR 1822-4770 Foil.jpg
GR 1822-4770 Foil.jpg|1822-4770
GR 1822-4776 Component.jpg
GR 1822-4776 Component.jpg|1822-4776
GR 1822-4776 Foil.jpg
GR 1822-4776 Foil.jpg|1822-4776
GR 1822-4780 Component.jpg
GR 1822-4780 Component.jpg|1822-4780
GR 1822-4780 Foil.jpg
GR 1822-4780 Foil.jpg|1822-4780
GR 1822-4785 Component.jpg
GR 1822-4785 Component.jpg|1822-4785
GR 1822-4785 Foil.jpg
GR 1822-4785 Foil.jpg|1822-4785
GR 1822-4790 Component.jpg
GR 1822-4790 Component.jpg|1822-4790
GR 1822-4790 Foil.jpg
GR 1822-4790 Foil.jpg|1822-4790
GR 1822-4796 Component.jpg
GR 1822-4796 Component.jpg|1822-4796
GR 1822-4796 Foil.jpg
GR 1822-4796 Foil.jpg|1822-4796
</gallery>
</gallery>



Latest revision as of 01:58, 4 July 2024

General Radio 1822
Digital Voltmeter Calibrator
General Radio 1822 Digital Voltmeter Calibrator

Available from 1970 to 1970

Manuals
  • Manual Needed.
Catalog History
 Document Year Page
Catalog U 1970 257
(All manuals in PDF format unless noted otherwise)

The General Radio 1822 Digital Voltmeter Calibrator was available in Catalog U (1970) only.

The Type 1822 is designed to produce accurate voltages for the testing and calibration of digital voltmeters. It can be used as a precise laboratory voltage reference from 100 μV DC to 1111.1 V DC. The output voltage is displayed using a mechanical digital display with automatic units and decimal placement. It can set to any value with five-digit resolution, the resolution is adjustable.

External programming is available through a rear-panel connector along with an internal automatic step function. Other features include an interference voltage function, polarity reversal, and adjustable output impedance.

Specifications

  • Output Voltage: 100 μV DC to 1111.1 V DC in 7 decade ranges
  • Output Impedance: <0.5 Ω on 1 V through 1 kV ranges; 200 Ω on 1 mV through 100 mV ranges; A 1 kΩ, 10 kΩ, or 100 kΩ output resistor can be added to above impedances by front-panel selector switch
  • Output Current: 10 mA at full range on 1, 10, and 100-V ranges; 1 mA on 1 kV range
  • Accuracy: ±(30 ppm of voltage + 10 ppm of full range + 3 μV) on all ranges except ±(50 ppm of voltage + 10 ppm of full range) on 1-kV range; Accuracy of rear-panel voltages ≤111 mV, ±(30 ppm of voltage + 10 ppm of range +0.5 μV)